-40%

AMETEK CRYSTAL IS33-36 30 SERIES 3000PSI PRESSURE CALIBRATOR GAUGE

$ 1002.66

Availability: 100 in stock
  • All returns accepted: Returns Accepted
  • Refund will be given as: Money Back
  • Restocking Fee: No
  • Return shipping will be paid by: Buyer
  • Condition: New
  • Item must be returned within: 30 Days
  • Product: Pressure Calibrator
  • Brand: CRYSTAL
  • MPN: iS33-36/3000PSI

    Description

    NEW - NEVER PUT INTO SERVICE
    AMETEK CRYSTAL IS33-36
    30 SERIES 3000PSI PRESSURE CALIBRATOR GAUGE
    ITEM INFORMATION
    ITEM:
    PRESSURE CALIBRATOR GAUGE
    MAKE:
    AMETEK CRYSTAL
    MODEL:
    IS33-36/3000 PSI
    NEW/USED:
    NEW - PURCHASED FOR PLANT EXPANSION THAT WAS CANCELLED - NEVER PUT INTO SERVICE
    FUNCTIONALITY:
    NEW - PURCHASED FOR PLANT EXPANSION THAT WAS CANCELLED - NEVER PUT INTO SERVICE - 30 DAY WARRANTY
    COSMETIC:
    SEE PHOTOS
    ACCESSORIES:
    NEW - NEVER PUT INTO SERVICE - AS SEEN - 30 DAY WARRANTY
    ITEM NUMBER:
    R7S1 - 2262-822715, 2262-822718
    Crystal iS33 Pressure Calibrator
    The 30 Series are compact and rugged pressure calibrators designed to bring laboratory accuracy to outdoor field conditions. Every 30 Series Calibrator is fully temperature compensated through extensive testing of all measured parameters while being exposed to environments of 0°C (32°F) to 50°C (122°F).
    All 30 Series calibrators are intended for gauge pressure measurement. That is, they indicate the difference between applied pressure and ambient barometric pressure. The zero button also can be used as a tare function - extending the functionality of the calibrator for special applications.
    They weigh less than a pound and advanced technology is employed throughout the product line. Sensors are constructed with stainless steel isolation diaphragms and permanent oil fill. This means 30 Series products can be used with gases and liquids.
    Long battery life is achieved with a low power, RISC (reduced instruction set) type micro-controller for computation of complex algorithms. A 24 bit analog to digital converter provides internal resolution of 1 part in 16.7 million. Internal resolution always exceeds the displayed resolutions for all measurements. Full accuracy is maintained even while the low battery icon is flashing.
    The 30 Series is specified in percent of reading - instead of percent of scale. Why? Because one 30 Series calibrator could replace several conventional calibrators by covering a wide range of pressure with high accuracy.
    The Model 30 does not require any manual adjustments or trimming. All calibration is accomplished via an optically isolated communications port using a USB physical interface. The optical isolation protects the Model 30 from electrical transients that may appear on the USB connector or cabling to a PC.
    Your 30 Series can be customized through the use of our free ConfigM30ル software. Your personal computer can disable, enable, or modify a variety of features of your 30 Series.
    Features:
    Add or create new pressure units and/or disable unused pressure units
    Password protection to prevent unauthorized changes
    Load and save custom configurations
    View and print nan as received versus as left change report
    Select your preferred H20 water density: 4°C, 60°F, or 68°F/20°C
    Store a message or identification number