-40%

GE MAC5500 ECG Machine

$ 1399.2

Availability: 100 in stock
  • Condition: Seller refurbished
  • All returns accepted: ReturnsNotAccepted

    Description

    Built on GE’s proven innovation in ECG acquisition and analysis, the MAC® 5500 is GE’s premier ECG system, delivering advanced disease management capabilities through its industry-leading collection of standards-setting algorithms and advanced networking.
    The MAC 5500 system offers the sophistication required for advanced ECG applications, while its ease of use extends this level of performance to the broadest range of users possible. And, it’s part of the complete GE suite of networked, non-invasive testing solutions designed to maximize patient throughput and department productivity.
    • Advanced algorithms set the standard in ECG analysis and interpretation.
    • Easy-to-use applications and features streamline productivity and workflow.
    • Seamless connectivity to the MUSE® cardiology information system speeds data storage and ECG retrieval.
    Microprocessor augmented automatic electrocardiograph; 14-leadwire acquisition with programmable lead configuration
    Processing:
    ECG Interpretation:
    Marquette 12SL ECG Analysis Program for Adults and Pediatrics
    Computerized Measurements: 15-lead analysis includes measurements of user-selectable
    additional 3 leads
    Optional:
    Additional ECG Function: ECG Analysis Frequency: ECG Storage:
    External Archiving: Digital Sampling Rate: Pre-Acquisition:
    Hi-Res Late Potential Analysis and P-Wave Signal – Averaged ECG
    Vectorcardiography
    500 samples/second (sps)
    200 ECGs in internal memory Secure Digital card
    4,000 samples/second/channel
    Provides 10 seconds of instantaneous ECG acquisition
    Dynamic Range: AC Differential ± 5mV, DC offset ±320 mV
    Resolution:
    Frequency Response: Common Mode Rejection:
    4.88 μV/LSB @ 250 sps, 4.88 μV/LSB @ 500 sps
    –3 dB @ 0.01 to 150 Hz
    >140 dB (123 dB with AC filter disabled)
    Input Impedance: >10MΩ @ 10 Hz, defibrillator protected
    Patient Leakage: <10 μA
    Pace Detect: Orthogonal LA, LL, and V6; 750μV@50μs
    Special Acquisition Functions: Disconnected lead detection, electrode impedance, excessive,
    AC noise, baseline wander and muscle tremor messages
    Heart Rate Meter: 30 to 300 BPM ±10% or 5 BPM, whichever is greater. Heart rates
    outside this range will not be displayed.
    Includes: Cam 14 Module with Leadwires and Patient Cable, Paper, Universal Electrode Clips, Electrodes,
    Battery, Power Cord, Biomed Certificate